Simultaneous in vivo magnetic resonance imaging and radioactive measurements with the β-MicroProbe
Titre de la revue : European Journal of Nuclear Medicine and Molecular Imaging Volume : 34 N° : 11 Pagination : 1868-1872 Date de publication : 01/11/2007
Purpose: Multimodal instrumentation is a new technical approach allowing simultaneous and complementary in vivo recordings of complementary biological parameters. To elucidate further the physiopathological mechanisms in intact small animal models, especially for brain studies, a challenging issue is the actual coupling of magnetic resonance imaging (MRI) techniques with positron emission tomography (PET): it has been shown that running the technology for radioactive imaging in a magnet alters the spatiotemporal performance of both modalities. Thus, we propose an alternative coupling of techniques that uses the β-MicroProbe instead of PET for local measurements of radioactivity coupled with MRI. Methods: We simultaneously recorded local radioactivity due to [18F]MPPF (a 5-HT1A receptor PET radiotracer) binding in the hippocampus with the β-MicroProbe and carried out anatomical MRI in the same anaesthetised rat. Results: The comparison of [18F]MPPF kinetics obtained from animals in a magnet with kinetics from a control group outside the magnet allowed us to determine the stability of tracer biokinetic measurements over time in the magnet. We were thus able to show that the β-MicroProbe reliably measures radioactivity in rat brains under an intense magnetic field of 7 Tesla. Conclusion: The biological validation of a β-MicroProbe/MRI dual system reported here opens up a wide range of future multimodal approaches for functional and pharmacological measurements by the probe combined with various